Evaluation of the clinical validation and efficiency of on-line urea monitor
LI li ,LI Xuewang ,LIU Hong,et al.
(Division of Nephrology,Peking Union Medical College Hospital,Beijing 100730,China)
Abstract:Objective To evaluate the clinical validation and efficiency of on-line urea monitor.Methods Quantitative parameters (Kt/V,nPCR,REM,SRI and V) were evaluated with on-line urea monitor in 17 patients treated by hemodialysis.Results The paired analysis of urea removed (REM),as estimated by the urea monitor compared with the direct dialysis quantification (DDQ),showed no significant difference.Similarly,compared with urea kinetic model,measurements of Kt/V and nPCR by urea monitor showed no significant difference.When assessed over 1 week in a subgroup of 12 patients,Kt/V and nPCR remained relatively stable during 3 consecutive assessments,but urea removed and predialysis urea concentrations declined.Conclusion Compared with urea kinetic model,on-line urea monitor can give more accurate parameters of dialysis adequacy.
Keywords:Hemodialysis,utilizing;Kinetics,urea;Dialysate;Protein catabolic rate
参考文献:
[1]Lowrie EG,Laird NM.The National Dialysis Cooperative Study.Kidney Int,1983,23 Suppl 13:S1-S122.
[2]Daugirdas JT.Second generation logarithmic estimates of single-pool variable volume kt/V:an analysis of error.J Am Soc Nephrol,1993,4:1199-1204.
[3]Garred LJ,Bariehello DL,Canaud BC.et al.Simple equations for protein catabolic rate determination from pre dialysis and post dialysis blood urea nitrogen.ASAIO,1995,41:889-895.
[4]Watson PE,Watson ID,Batt RD.Total body water volumes for adult males and females estimated from simple antropometric measurements.Am J Clin Nutr,1980,33:27-39.
[5]Bosticardo GM,Avalle U,Giacehino F,et al.Accuracy of an on-line urea monitor compared with urea kinetic model and direct dialysis quantification.ASAIO,1994,40:M426-M431.
[6]Depner TA,Deshaviah PR,Ebben JP,et al.Multicenter clinical validation of an on-line monitor of dialysis adequacy.J Am Soc Nephrol,1996,7:464-471.
[7]Canaud B,Bose JY,Leblanc M,et al.Evaluation of high-flux hemodilafiltration efficiency using an on-line urea monitor.Am J Kidney Dis,1998,31:74-80.
收稿日期:2000-03-21