中图分类号:R318 .6文献标识码:B文章编号:1005-202X(1999)03-0159-02
The development of oxygen therapy device with constant temperature and proper humidity
LI Guang-zhong
Physical department,LIU Jun-ying2
Internal medicine department,Binzhou medical college, Shandong 256603,China
Abstract:To introduce a new oxygen therapy device which can provide moderate temperature and humidity.It consists of humidifier, controlling circuits of temperature and lowest water level, pipelines of oxygen therapy. It has resolved the problems of respiratory tract dryness, ropy sputum, dead cavity increase resulting from the traditional oxygen therapy device. Besides, this device has the following characteristics: easy modulation and operation, a small volume, dependable safety.
Key words:constant temperature ; oxygen therapy device ; humidifier
引言:吸氧装置是抢救危重病人的必备装置之一,在抢救呼吸系统危重病人时尤其重要。理想的吸氧装置要求病人吸入的氧气除保持适当的浓度和流速外,还要有合适的温度和湿度,一般吸入氧气的温度要保持在36 ℃左右,湿度在80%左右,近于生理上的要求。但目前临床上通用的吸氧装置无恒温和加湿功能,易于造成病人呼吸道干燥,导致痰液粘稠、死腔增多,从而增加了病人痛苦。为此,我们设计了一种自动恒温、加湿吸氧装置,有效的解决了以上难题。该装置体积小、重量轻、易于调节、使用方便,安全可靠,是一种经济实用的医疗器械。
1结构及工作原理
恒温加湿吸氧装置主要由湿化器、温度控制电路、最低水位监测报警电路、输氧管道和吸氧面罩等组成,原理如图1所示。
图1原理框图
1.1输氧管道、吸氧面罩
氧气通过湿化器与热蒸气混合后经输氧管道送至吸氧面罩,温度传感器安装在吸氧面罩内,用以检测混合气体的温度。
1.2湿化器、温度控制电路
湿化器由湿化瓶、100 W电热加热器构成;220 V交流电经隔离变压器给加热器供电,对湿化瓶内的液体进行加热;温度控制电路主要由温度传感器、电压比较器、驱动电路和继电器等组成,如图2所示,由温度传感器检测气体温度后,经电压比较器与设定值进行比较,从而通过继电器J控制加热器电源的通断,将面罩内气体温度控制在 36 ℃左右。
图2电原理图
1.3最低水位监测报警电路
最低水位控制报警电路主要由水位传感器、电压比较器、驱动电路和蜂鸣报警器等组成,如图2所示,由水位传感器监测湿化瓶内水位的变化,当水位低于设定的最低限时,电压比较器发生翻转,蜂鸣报警器发出缺水报警,同时通过T1、T2控制继电器J断开加热器电源,以确保加热器不因湿化瓶被烧干而损坏。
2性能和特点
(1)装置可实现湿化和自动恒温的功能,对吸氧易于造成的呼吸道干燥、痰液粘稠、死腔增多等不适症有明显的改善,可有效地减轻输氧病人的痛苦,提高治疗效果。
(2)装置设计简单,制作方便,加热电源与市电通过隔离变压器实现了有效隔离,安全可靠。
(3)装置具有体积小、重量轻、易于调节、使用方便等特点,是一种经济实用的医疗器械,可广泛用于各级医疗卫生部门。
参考文献:
[1] 汪丽蕙,等。今日内科(呼吸疾病分册)[M].北京:北京医科大学、中国协和医科大学联合出版社,1992.
收稿日期:1998-11-11